Tuttnauer Vertical Sterilizer (160 Liters)
Details
- Independent temperature and pressure monitoring
- Chamber constructed of 316Ti grade stainless steel
- Accelerated cooling technology reduces cycle time
- 23 to 160 Liter autoclave chamber volume
- Sterilizes liquids, solids, glassware and biohazard
- Advanced microprocessor control panel
- STERILIZATION APPLICATIONS
- Liquids such as nutrient media and buffer solutions
- Solid items such as pipettes, tubes and filters
- Glassware and plastic articles
- Biohazard sterilization
- The various applications can be carried out using the autoclave’s standard programs.
- Additional options ensure more control over results and optimal, safe and repeatable sterilization processes.
- For special applications, our experienced team will customize a suitable method for you.
- CONTROL AND DOCUMENTATION
- Control Features
- High precision control system for perfect sterilization results
- Password protection allows for secure access control
- Independent temperature and pressure monitoring
- Cycle information recovery in the case of power failure or cycle interruptionFail Alert – Indicates cycle failure or interruptionDoor Alert – Indicates the door is unlocked
- RS 232 PC Connection Port for direct software updates and remote maintenance
- Built in Printer
- The laboratory line can be supplied with a built-in printer to document the performed cycle, in addition to a comprehensive LCD display. This feature is optional.
- ADMC Software
- The ADMC software allows complete control and monitoring of up to eight sterilizers.
- The software retrieves data, creates graphs, tables and print outs, controls the autoclave and shows real-time data.
- SAFETY FEATURES
- Door Safety
- The autoclave lids are designed with a number of independent mechanical and digital safety features.
- A safety device prevents the operator from opening the door when the chamber is pressurized
- Steam is not allowed to enter the chamber when the door is open
- A cycle cannot start if the door is open or not properly locked
- The door cannot unlock until liquid temperature reaches the predetermined end temp
- The door cannot unlock until chamber pressure reaches room pressure
- Temperature Activated Door Lock - The door will not open until the temperature is below a specified safe level
- General Safety Features
- Double Independent Monitoring:
- The combined electronic and mechanical monitoring guarantees accurate results. The operator has two independent means to monitor temperature and pressure.
- Safety Valves:
- The chamber is equipped with safety valves – if the pressure exceeds the allowed limit the safety valves will discharge
- Built-in Steam Generator Safety:
- A water level monitoring system maintains a constant water level and ensures safe operation of the heaters
- QUALITY FEATURES
- The chamber is constructed of long lasting 316Ti grade stainless steel with superior corrosion resistance
- The generator is made from stainless steel
- The autoclave automatically switches to standby mode if no buttons or switches are operated for four hours
- The autoclave is designed for easy servicing allowing maintenance access to all components
- Drain Protection: The exhaust drain is mixed with cold tap water that cools the exhaust’s temperature before reaching the drain
- STERILIZER DIRECTIVES AND STANDARDS COMPLIANCE
- Tuttnauer’s high quality laboratory autoclaves comply with the strictest international directives and standards.
- Pressure Equipment:
- PED 97/23 EEC, EN 10028-7,ASME Code Sec VIII
- Safety:
- IEC/UL/EN61010-1, IEC 61010-2-040, EN 61326
- Sterilization:
- DIN 58951 Series, ISO 17665-1:2006
- Quality System:
- ISO 9001:2000, ISO 13485:200