Tuttnauer Pre & Post Vacuum Sterilizers 160 Liters

Item #: Z-MAR-5075-HSG
$35,700.00

Details

The vacuum autoclave features wheels for easy mobility and sterilization on demand.

  • Stainless Steel Chamber and Door with Electro-Polish Finish
  • High-Volume Water-Ring Vacuum Pump
  • Standby Heating Mode
  • Built-in Alphanumeric Printer
  • Complete Air Removal

A pre-vacuum pump removes the air from the chamber and then admits a pulse of steam

The unit includes a built-in alphanumeric printer. Unit requirements include a 5 cm (2") building drain (floor sink recommended) and 4750 BTU/hour heat load to room. Additional features include two RTD load probes for liquid cycles and a double-locking safety device that prevents the door from opening at high pressure and high temperature. The control lock-out switch prevents the start of a cycle if the door is not properly locked. The vacuum autoclave includes a 316-grade stainless steel chamber and door with electro-polish finish and a high-volume water-ring vacuum pump for fast and efficient air removal. The standby heating mode keeps the autoclave warm and ready for use.

This process repeats one-to-four times according to selection. With each cycle of vacuum and steam-pulse, the air fraction decreases. This allows for complete air removal and faster and more efficient steam penetration throughout the entire load.


Chamber Dimensions (Ø x L)
19.5" x 29.5"
Chamber Volume
160 Liter
Number of Shelves
2
Shelf Dimensions (WxHxL)
Small: 11.9" x 0.8" x 27.6"
Large: 18.6" x 0.8" x 27.6"
Electrical Requirements (Control System, Vacuum Pump)
1-Phase, 120V, 20 AMP
Electrical Requirements (20kW Generator)
3-Phase, 208V, 56 AMP
or
3-Phase, 480V, 25 AMP
Cold Water Supply
3/4" NPT
Overall Dimensions (WxHxL)
33.6” x 65.4” x 44.8”
Building Drain
2"
*Floor Sink Recommended
Heat Load to Room
4750 Btu/Hr
Autoclave Weight
935 Lbs