Pelton and Crane OCM Sterilizer - Refurb
Details
- This is a work horse of a sterilizer. It will work forever for you. Pelton & Crane OCM Sterilizer. Stainless steel door is secured at four points and incorporates a door interlock switch to ensure that door is properly locked before cycle begins.
- This sterilizer was upgraded to shut off and vent automatically at the end of the cycle.
- It takes about 20 minutes to get to 273f, and then the general guidelines:
- General Guidelines*
- Unwrapped instruments..........121c 250f 15psi 15min
- 132c 270f 30psi 3min
- Lightly wrapped.......................121c 250f 15psi 20min
- 132c 270f 30psi 9min
- Heavily wrapped...................... 121c 250f 15psi 25min
- 132c 270f 30psi 11min
- Chamber size is 7" dia x 14" deep.
- Outside dimensions 15" W x 18" D x 13" H
- 110V
- This is a work horse of a sterilizer. It will work forever for you. Pelton & Crane OCM Sterilizer. Stainless steel door is secured at four points and incorporates a door interlock switch to ensure that door is properly locked before cycle begins.
Here is what is being done to this sterilizer before it is being shipped to you:
Replace all three heaters
Replace filter
Install New Gasket
Rewire the entire sterilizer with temperature sensitive wire
Upgrade to a larger reservoir
Install New Temperature and Pressure Gauges
Calibrate for more accurate temperature & pressure reading
Install 'Over Pressure' Safety Valve
Upgrade to our Automatic Shut Off System
Upgrade to our Automatic Venting System
Upgrade and strengthen the chamber.This is a Refurbished Sterilizer